Micro leak of the cuff inflation system of an armored endotracheal tube: an unreported site

To the Editor,

A 44-year-old male posted for lumbar (L4–5) microdiscectomy was intubated with 7.5-mm internal diameter cuffed armored endotracheal tube (cAETT) for general anesthesia after routine preoperative check of the cuff inflation system. Anesthesia was maintained with O2, air, and sevoflurane and was positioned prone. He was ventilated in volume control mode with tidal volume 450 ml with cuff pressure of 25 cmH2O. His airway pressure was 22 cmH2O. His expired tidal volume (Vte) decreased to 350 ml, and cuff pressure was only 15 cmH2O, 1.5 h after intubation and 1 h after prone positioning. The cuff was reinflated to 25 cmH2O, and the set tidal volume (450 ml) got delivered for the rest of the surgery (1 h) and extubation was uneventful. Post extubation, the methylene blue injected through the cuff inflation system revealed a micro leak, where the cuff is attached to the outer wall of the cAETT. The other parts of the inflation system were perfectly sealed, and there was no leak (Fig. 1). Though many sites for leak of ETT cuff system were reported (El-Orbany and Salem 2013; Malhotra and Singhal 2006), this defect at the sealing of cuff to the ETT had not been reported. Though it was a manufacturing defect, it had been easily missed in the preoperative checking of the cAETT because of very slow leakage of air. Decrease in Vte with decrease in cuff pressure after 1.5 h of intubation in an undisturbed head and neck position suggests that it could be a micro leak from the cuff, when all other causes are ruled out. Such concealed micro leaks can be managed by reinflation of the cAETT and continuous monitoring of cuff pressure (Mariappan et al. 2015), instead of changing the ETT especially in prone position in the middle of surgery.

Fig. 1
figure 1

Micro leak of the cuff inflation system of an armored endotracheal tube: an unreported site. The methylene blue injected through the cuff inflation system revealed a micro leak, where the cuff is attached to the outer wall of the cAETT. The other parts of the inflation system were perfectly sealed and there was no leak

Informed written consent has been obtained from the patient for publication.

Availability of data and materials

Yes, they are available.

Abbreviations

cAETT:

Cuffed armored endotracheal tube

L:

Lumbar

Vte:

Expired tidal volume

References

  • El-Orbany M, Salem MR (2013) Endotracheal tube cuff leaks: causes, consequences, and management. Anesth Analg 117:428–434

    Article  Google Scholar 

  • Malhotra N, Singhal S (2006) Slow deflation of a tracheal tube cuff: an unusual cause. Acta Anaesthesiol Scand 50:519–520

    Article  CAS  Google Scholar 

  • Mariappan R, Harshit CR, Prasanna RG (2015) Novel method of identifying intraoperative cuff leak and its treatment while monitoring cuff pressure. J Neurosurg Anesthesiol 27:349–350

    Article  Google Scholar 

Download references

Acknowledgements

The authors acknowledged Mr. Peter Durai and Mrs. Rajakumari Pugazhenthi for their technical support.

Funding

No funding received.

Author information

Authors and Affiliations

Authors

Contributions

VV was involved in the clinical care of the patient, identifying the cuff leak, troubleshooting the leak, further safe management of the patient, identifying the cause for the leak, reviewing the literature, and drafting the manuscript. GA was involved in troubleshooting the leak, safe management of the patient, identifying the cause for the leak, reviewing the literature and drafting the manuscript. Both authors read and approved the final manuscript.

Corresponding author

Correspondence to Vinodhadevi Vijayakumar.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Informed written consent has been obtained from patient for publication.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Vijayakumar, V., Ganesamoorthi, A. Micro leak of the cuff inflation system of an armored endotracheal tube: an unreported site. Ain-Shams J Anesthesiol 11, 27 (2019). https://doi.org/10.1186/s42077-019-0042-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s42077-019-0042-x